Recall Z-2122-2014 ยท FNL Bed, Ac-Powered Adjustable Hospital

Stryker Medical Division of Stryker Corporation

InTouch Critical Care Bed, Stryker Medical. Model Numbers 2130, 2131, 2140, 2141, 2151, 2152, 2155, and 2156.

DateJul 28, 2014 (initiated Jun 4, 2014)
StatusTerminated
ReasonStryker has received complaints from the field alleging instances where the In Touch beds are hoisted or lifted from the floor and the casters have fallen off. The investigations into these casters have shown that in these cases the caster has broken on the top of the caster stem where the brake rod inserts into the caster. If more than one caster on a unit is damaged, brake functions can be affec
Root causeDevice Design
ActionOn 06/30/1014 the firm sent "Urgent Medical Device Notification" letters to their customers.
Quantity11,632 units
DistributionWorldwide Distribution -USA (nationwide) including DC and the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI & WV, and the countries of Argentina, Bra