Recall Z-2002-2014 ยท FRC Indicator, Biological Sterilization Process

3M Company - Health Care Business

3M Attest Auto-reader. Model numbers 390, 390G,and 490. These devices incubate the biological indicators (BIs) to determine a positive or negative result for the sterilization cycle.

DateJul 7, 2014 (initiated Jun 9, 2014)
StatusTerminated
ReasonLabeling on units shipped prior to May 16th, 2014 did not contain the statement "This product contains dry natural rubber" as required by the United States Food and Drug Administration (FDA) for medical devices. The non-slip pads at the bottom of the unit contain dry natural rubber. One incident of an allergic reaction in a sensitized individual has been reported.
Root causeLabeling Change Control
ActionConsignees were sent on 6/9/2014 a 3M "Urgent Medical Devices Correction" letter dated May 21, 2014. The letter described the problem and the product involved in the recall. The letter described what action is 3M taking and what action was required of the consignees. Consignees were sent an adhesive label and a response card. They requested consignees to adhere the label to a portion of the plastic covering each unit in a location where it is visible to the end-user. After that, they are requested to fill out and return the response card.
Quantity5343 (2621 USA, 2722 OUS)
DistributionWorldwide Distribution-USA (nationwide) including DC and Puerto Rico and the countries of AUSTRALIA, CANADA, CHILE , COLOMBIA, COSTA RICA, SOUTH AFRICA, BRAZIL, GERMANY, ECUADOR, DUBAI, JAPAN, HONG KONG , CHINA, KOREA, MALAYSIA, PANAMA, SINGAPORE, VATICAN, and THAILAND.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K1032773M ATTEST 1491 SUPER RAPID READOUT BIOLOGICAL INDICATOR; 3M ATTEST 490 AUTO-READER
3M Company-3M Health Care, Apr 19, 2011
K1235463M ATTEST(TM) AUTO-READER
3M Company, Mar 18, 2013