Recall Z-1872-2014 ยท HRX Arthroscope

Stryker Endoscopy

Shaver Handpiece Sterilization Tray. Model number 272-700-000. For sterilization of Stryker Endoscopy arthroscopic shaver.

DateJun 25, 2014 (initiated May 19, 2014)
StatusTerminated
ReasonSTERRAD 100S parameters provided in the Shaver Handpiece reprocessing guide (P10299 Revision A) did not fully consider all worst-case scenarios during validation testing.
Root causeNonconforming Material/Component
ActionDevice Correction letters were sent on May 23, 2014 to all users by certified mail. Internation sites will receive notification and be responsible for contacting customers.
Quantity1992 trays
DistributionWorldwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K941333SE5 ARTHROSCOPY SYTEM
Stryker Endoscopy, Nov 10, 1994