Recall Z-2075-2014 ยท MUJ System, Planning, Radiation Therapy Treatment

Siemens Medical Solutions USA, Inc

Siemens syngo RT Dosimetrist 2.7 system Product Usage: The intended use of syngo Suite for Oncology Systems is as an accessory to the linear accelerator system to aid and support in the planning and delivery of x-ray radiation for the therapeutic treatment of cancer.

DateJul 17, 2014 (initiated May 15, 2014)
StatusTerminated
ReasonAn update to the Virtual Simulation software of the syngo RT Dosimetrist is necessary to address a safety issue. In special scenarios within the Structure Operations function the system did not store the selected name from the list of predefined names. After reloading of the data, this structure name had been changed by the system. Additionally, a performance issue was addressed where import of DI
Root causeSoftware design
ActionSiemens sent a Medical Device Correction Field Safety Notice of Improvement letter to affected customers via USPS Certified Mail. The letter described the affected product, problem, and actions to be taken.
Quantity7
DistributionUS Nationwide Distribution in the states of NH, TN, OH, VA, and TX.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K101119SYNGO DOSIMETRIST WORKSPACE V2.7 MODEL: 1056864
Siemens Medical Solutions USA, Inc., Jun 16, 2010