Recall Z-1824-2014 ยท DQA Oximeter

Spacelabs Healthcare Inc

Spacelabs Healthcare Ventilator Flexport Interface, Model 90436A-07

DateJun 18, 2014 (initiated May 9, 2014)
StatusTerminated
ReasonSpacelabs Healthcare is voluntarily recalling the Hamilton Galileo Ventilator Flexport, Model 90436A-07, where the monitored Minute Volumes (Vmin) has been reported at one time to reach ten times the actual value on the bedside monitor.
Root causeComponent change control
ActionSpacelabs changed their strategy for updating the software and sent the updated letter Urgent-Medical Device Correction- Hamilton Galileo Ventilator Flexport Model 90436A-07, dated 10 June 2015, to their consignees. The firm is going to send consignees new replacement flexport(s) with upgraded software and requested consignees to return old affected units via Return Goods Authorization (RGA) to Spacelabs and the firm will destroy affected units. Consignees are instructed to take the following actions: 1. Please replace the affected devices and test for correct operation. 2. Return your ol
Quantity46 units distributed within the US; 108 units distributed outside the US.
DistributionWorldwide Distribution-USA (nationwide) including the states of AZ, IN, LA, GA, SC, FL, and NY, and 18 customers outside the US.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K903702MODEL 90421 FLEXPORT INTERFACE
Spacelabs, Inc., Oct 19, 1990