Recall Z-1800-2014 ยท IYE Accelerator, Linear, Medical
Elekta, Inc.
MOSAIQ, an oncology information system used to manage workflows for treatment planning and delivery. It supports information flow among healthcare facility personnel and can be used wherever radiotherapy and/or chemotherapy are prescribed.
| Date | Jun 16, 2014 (initiated May 23, 2014) |
| Status | Terminated |
| Reason | When using CMA, if the user hovers the mouse pointer exactly above the direction dropdowns and then uses the mouse thumbwheel or touchpad scroll bar, the direction selection might change when the intention was only to move the mouse to a different area of the screen. |
| Root cause | Software design |
| Action | Elekta sent an Important Safety Notice on May 23, 2014, to all affected customers. The notice identified the product the problem and the action needed to be taken by the customer. "Notify the affected customers of the issue - To be completed 06/22/2014 "Provide a fix to the defect in an expeditious timeframe. It is mandatory for all customers to apply the fix - Version 2.50.05 SP8 & Version 2.60 - To be released 05/16/2014 "Follow up with the affected customers for: -Confirmation receipt of the LIN, AND -Confirmation of installation of patch for the fix- To be completed 11/30/2014. For f |
| Quantity | 723 |
| Distribution | Worldwide Distribution - USA (nationwide) including Puerto Rico, and the countries of Albania, Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, E |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K123230 | MOSAIQ ONCOLOGY INFORMATION SYSTEM Impac Medical Systems, Inc., Dec 12, 2012 |