Recall Z-1775-2014 ยท HBE Drills, Burrs, Trephines & Accessories (Simple, Powered)
Stryker Instruments Div. of Stryker Corporation
Stryker 3.0mm x 3.8mm Neuro Drill Soft Touch Sterile Product Usage: The devices are intended to be used to cut bone and bone related tissue during neurosurgery, including craniotomy and spinal surgery as well as ear, nose and throat (ENT), orthopedic, and general surgical applications including maxillofacial, craniofacial surgeries. The main procedures/conditions in which these products are expec
| Date | Jun 11, 2014 (initiated May 22, 2014) |
| Status | Terminated |
| Reason | There may be missing notches, partial notches or no notches at all in some Cutting Accessories that are used with the Elite" Attachments. |
| Root cause | Nonconforming Material/Component |
| Action | Stryker Instruments sent an URGENT MEDICAL DEVICE RECALL notification letter dated May 22, 2014, to the consignees. Recall notification included affected product description and identification, reason for recall, risk to health, and actions to be taken by the customers. Customers will be instructed to review the notification, quarantine unused product and to return a signed Business Reply Form. They will be advised that a Stryker Representative will be sent to their facility to inspect any affected product on hand. Once product is inspected by the Stryker Representative, they can return co |
| Quantity | 802 |
| Distribution | Worldwide Distribution - US (nationwide) |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.