Recall Z-1746-2014 ยท MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)

Philips Medical Systems, Inc.

IntelliVue MX40 Wearable Patient Monitor in use with the Philips IntelliVue Information Center iX (PIIC iX); 865350, 865351, 865352

DateJun 10, 2014 (initiated May 22, 2014)
StatusTerminated
ReasonIntelliVue MX40 (part numbers 865350, 865352) with software revisions B.00 or B.01 may not resume alarming when the alarms are paused and the pause timer ends.
Root causeSoftware design
ActionOn May 22, 2014, each affected customer was sent an Urgent Medical Device Correction notification/Field Safety Notice. The letter explains the problem, explains the action to be taken by the consumer, and the action planned by Philips. The Urgent Medical Device Correction notification/Field Safety Notice informs customers of the issue, identifies details of the units affected, gives instructions on actions to be taken by the customer and identifies what action Philips plans to take to remedy the issue. The correction will consist of a free of charge software upgrade. A Philips Healthcare
Quantity42,600
DistributionUnited States: Nationwide Foreign Countries: Germany, Austria, Finland, Sweden, Australia, Qatar, Canada, Netherlands, New Zealand, Portugal, United Kingdom, Spain, Belgium, France, Switzerland, Norway, Czech Republic, Poland, Italy, Singapore, Denmark, Iceland, India, Saudi Arabia, Ireland, Oman,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K103646INTELLIVUE MX40 PATIENT MONITOR
Phillips Medical Systems, Jan 7, 2011
K113125INTELLIVUE MX40 PATIENT MONITOR
Philips Medical Systems, Nov 17, 2011