Recall Z-1958-2014 ยท OCC Respiratory Virus Panel Nucleic Acid Assay System

Focus Diagnostics Inc

Simplexa" Flu A/B & RSV Direct, Model Number: MOL2650. 510(k) K120413 Simplexa" Flu A/B & RSV Direct. The Focus Diagnostics Simplexa Flu A/B & RSV Direct assay is intended for use on the 3M Integrated Cycler instrument for the in vitro qualitative detection and differentiation of influenza A virus, influenza B virus, and respiratory syncytial virus (RSV) RNA in nasopharyngeal swabs (NPS) fro

DateJul 1, 2014 (initiated May 15, 2014)
StatusTerminated
ReasonFocus Diagnostics is providing an urgent safety notice for a correction to the labeling for Simplexa" Flu A/B & RSV Direct (MOL2650). Focus Diagnostics received some customer complaints of Simplexa Flu A/B & RSV Direct assays with sporadic false signals, which may result in a higher false results rate due to a potential software spectral matrix and Direct Amplification Disc storage.
Root causeDevice Design
ActionFocus Diagnostics sent an Urgent Safety Notice Correction letter dated May 15, 2014, to all affected customers. The notification, informed customers of the product, the problem, and the action to be taken by the customer. To decrease the likelihood of obtaining any false results caused by either manually entering a spectral matrix or storing the DAD improperly after each use, please follow the recommended warnings and precautions provided in the revised and attached Simplexa Flu A/B & RSV Direct (PI.MOL2650 Rev. D) package insert. 1. The spectral matrix must be installed in each 3M Integrate
Quantity614 kits
DistributionNationwide Distribution
FDA recordOpen on fda.gov