ClearanceBrief
Recall Z-2068-2014 ยท CCK Analyzer, Gas, Carbon-Dioxide, Gaseous-Phase

GE Healthcare, LLC

DATEX-OHMEDA S/5 SINGLE-WIDTH AIRWAY MODULE, E-MINIC AND ACCESSORIES. DATEX-OHMEDA S/5TM FM WITH L-FICU04 ANDL-FICUO4A SOFTWARE AND N-FCREC MODULE **(OUS only). E-miniC and accessories is indicated for monitoring C02 and respiration rate of all hospital patients.

DateJul 18, 2014 (initiated Jun 11, 2014)
StatusTerminated
ReasonFailure of the CO2 detector in Single-width Airway and Extension modules, which may cause a slow continuous decrease of measured Et-/FiCO2 values. Incorrect Et CO2/Fi CO2 value may impair clinical decision making for both mechanically and spontaneously ventilated patients because of incorrect low CO2 values.
Root causeDevice Design
ActionGE Healthcare sent an Urgent Medical Device Correction letter dated June 11, 2014, to all affected consignees. The letter described the Safety Issue, Safety Instructions, Affected Product Details, and Product Correction and Contact Information. For questions customers were instructed to contact Technical Support at 1-800-558-7044 or their local Service Representative.
Quantity2,948 (761 units US, 2,187 units OUS)
DistributionWorldwide Distribution - USA nationwide, ARGENTINA, AUSTRALIA, AUSTRIA, BAHRAIN, BRAZIL, CANADA, CHILE, CHINA, Croatia, CZECH REPUBLIC, DENMARK, EGYPT, EL SALVADOR, FINLAND, FRANCE, GERMANY, GREECE, GUATEMALA, HONG KONG, INDIA, IRAQ, ISRAEL, ITALY, JAPAN, KAZAKHSTAN, REPUBLIC OF KOREA, MALAYSIA, NET
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K043276DATEX-OHMEDA S/5TM FM WITH L-FICU04 AND L-FICU04A SOFTWARE AND N-FCREC MODULE
Datex-Ohmeda, Dec 23, 2004
K052582DATEX-OHMEDA S/5 SINGLE-WIDTH AIRWAY MODULE, E-MINIC AND ACCESSORIES
GE Healthcare, Oct 14, 2005