Recall Z-1842-2014 ยท LPH Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented

Zimmer, Inc.

VerSys Hip System, Beaded Fullcoat Stems, Zimmer, Inc., Warsaw, IN 46581

DateJun 20, 2014 (initiated May 19, 2014)
StatusTerminated
ReasonDuring routine inspection, beaded hip stem forgings received from supplier failed to meet ZES 2A-102. Processing of Beaded CoCrMo Alloy Hip Stem, material requirements per ATS #14-04818 and ATS #14-05549, respectively.
Root causeNonconforming Material/Component
ActionOn 6/2/2014 the firm sent "Urgent Medical Device Recall" letters to their customers. On 8/13/14, Zimmer contacted a distributor via phone pertaining to one additional lot (00-7843-011-06; lot 62547211) which had been included in the recall. Formal recall notices were issued to the distributor via email on 8/15/14.
Quantity147 units
DistributionWorldwide Distribution-USA (nationwide) including the states of AL, AZ, CA, CO, CT, FL, GA, IL, IN, MA, MI, MN, NC, ND, NJ, NY, OH, OK, PA, TN, TX, VA & WI, and the countries of Canada, Colombia, Dominican Republic, Japan, Korea (S. Korea), Peru, Taiwan & Thailand.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K973714VERSYS HIP SYSTEM BEADED HIP PROSTHESIS
Zimmer, Inc., Dec 24, 1997