Recall Z-1838-2014 ยท PBI Prosthesis, Hip, Constrained, Cemented Or Uncemented, Metal/Polymer, + Additive

Biomet, Inc.

G7 Acetabular System Positioning Guide Post, orthopedic surgical instrument for hip prosthesis. PN:110003500.

DateJun 19, 2014 (initiated May 7, 2014)
StatusTerminated
ReasonGuide rod may fracture during impaction of the acetabular shell and be retained in the inserter or the patient and could result in a delay in surgery.
Root causeUse error
ActionThe firm issued the following notice on May 7, 2014: URGENT MEDICAL DEVICE RECALL NOTICE: This notification is to inform you of an Urgent Medical Device Recall initiated by Biomet Orthopedics ("Biomet") which involves Part Numbers: 110003458 G7 Positioning Guide Rod and 110003500 G7 Positioning Guide Post. These instruments have been consigned and/or invoiced to your account. Biomet has initiated this action following an investigation which identified that positioning guide rod may fracture during impaction of the acetabula shell. If the positioning guide rod fractures, then the positioning g
Quantity1047 total
DistributionWorldwide Distribution-USA (nationwide) and the countries of Canada, Netherlands, New Zealand, and Japan.