Recall Z-2141-2016 ยท MEA Pump, Infusion, Pca
Hospira Inc.
LifeCare PCA infusion system with Hospira MedNet software allows clinicians to administer, or patients to self-administer analgesia, safely and effectively within clinician programmed limits and/or hospital-defined medication limits. The LifeCare PCA infusion system is used in a wide range of clinical settings that includes but is not limited to the following: medical, laboratory/delivery/post-par
| Date | Mar 8, 2013 |
| Status | Terminated |
| Reason | Hospira has received reports of PCA pumps not detecting distal occlusions, one of which resulted in a serious injury. The issue is caused by normal wear and tear on the Half Nut (the component/nut that travels up and down the lead screw) which prevents it from properly detecting the pressure build-up associated with a distal occlusion. |
| Root cause | Component design/selection |
| Action | Hospira sent an IMPORTANT DEVICE INFORMATION letter dated March 8, 2014 to all affected customers.. The letter advised customers to: 1) Perform the Performance Verification Test (PVT) Occlusion Test as defined in the PCA Technical Service Manual (TSM); 2) If the device does not pass the test, remove it from clinical service and contact Hospira at 800-441-4100 (Monday - Friday, 8:00 AM - 5:00 PM CST) to report the issue; and 3) Perform the appropriate troubleshooting and repair activities defined by your facility, which may include returning the device to Hospira for further diagnosing and serv |
| Quantity | 38,878 infusion pump systems |
| Distribution | Nationwide Distribution to AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY) + District of Columbia + Puerto Rico + Virgin Islands; *** FOREIGN: |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K042800 | LIFECARE PCA INFUSION SYSTEM WITH HOSPIRA MEDNET SOFTWARE Hospira, Inc., Oct 18, 2004 |