ClearanceBrief
Recall Z-1717-2014 ยท KPR System, X-Ray, Stationary

Siemens Medical Solutions USA, Inc

Siemens Ysio Systems Product Usage: The Ysio (New RAD-FAMILY) systems are radiographic systems used in hospitals, clinics, and medical practices. Ysio enables radiographic and tomographic exposures of the whole body including: skull, chest, abdomen, and extremities and may be used on pediatric, adult and bariatric patients. It can also be used for intravenous, small interventions (like biopsy,

DateJun 4, 2014 (initiated May 9, 2014)
StatusTerminated
ReasonExcessive usage of cleaning liquid on Siemens Ysio Systems may cause infiltration of the generator console, which may lead to a short circuit causing an unintended release of radiation.
Root causeUse error
ActionSiemens sent an Safety Advisory Notice letter dated May 9, 2014, to affected customer. The letter identified the affected product, problem and actions to be taken. For question call +49 (913) 84-9908.
Quantity433
DistributionUS Nationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K081722YSIO
Siemens Medical Solutions USA, Inc., Aug 22, 2008