Recall Z-1701-2014 ยท IZH System, X-Ray, Mammographic

Hologic, Inc.

Hologic M-IV Mammography System; Model M4--00001. Intended to produce radiographic images of the breast. For screening and diagnostic imaging of the breast.

DateJun 3, 2014 (initiated May 13, 2014)
StatusTerminated
ReasonSelenia Full Field Digital Mammography System defective relay switch, if there is a need to use the emergency off switch, the system may not turn off which could result in patient injury
Root causeComponent change control
ActionHologic notified customers by letter dated 5/13/14 via certified mail or Federal Express to inform customers of the problem and provide acknowledgement of the notification and confirmation that their systems have been corrected A Dealer notification dated 5/13/14 sent by certified mail or Federal Express to inform Dealers of the need to notify their customers of the problem and to correct the identified Selenia system using Hologic Technical Bulletin TB-00806. On 5/16/14 the firm expanded the recall to an additional units Mammography System M-IV (2 units) and Selenia Mammography Systems (4)
Quantity3 units
DistributionDistributed in Puerto Rico, and the states of CA, NY, and TX, and the country of Republic of Korea.