ClearanceBrief
Recall Z-1791-2014 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Roche Diagnostics Operations, Inc.

cobas 6000 analyzer series Operators Manual Roche cobas; Filter label: Filter (N) Mat. No 03149773001, Contents 1, 768-320 Filter (N) Product Usage: Fully automated immunoassay analyzer intended for the in-vitro quantitative/qualitative determination of analytes in body fluids.

DateJun 14, 2014 (initiated May 21, 2014)
StatusTerminated
ReasonUnder certain conditions loose ProCell/CleanCell aspiration tube filters may cause: consecutive discrepant results; sipper alarm; foam on reservoirs and system reagent alarms.
Root causeLabeling design
ActionRoche sent an Urgent Medical Device Correction (UMDC) letter dated May 20, 2014 via UPS Grount (receipt required) to affected customers. The letter identified the affected product, problem and actions to be taken. The letter instructed customers to: 1. Follow the instructions in the Precautions Regarding Handling the Aspiration Tube Filter section in this UMDC; 2. Discontinue performing ProCell/Clean Cell aspiration tube filter check as this maintenance item will be performed by your Roche FSR as a part of service maintenance; 3. If you have distributed the products listed within this UMDC t
Quantity1147
DistributionUS Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K060373COBAS 6000 SERIES SYSTEM
Roche Diagnostics, Mar 13, 2006