ClearanceBrief
Recall Z-1737-2014 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Beckman Coulter Inc.

Ultrasonic Transducer Kits For use with the Access Family of Immunoassay Systems*. *The Access Family of Immunoassay Systems includes the Access 2, UniCel DxI 800 and UniCel DxI 600, UniCel DxC 600i, and the UniCel DxC 880i, UniCel DxC 860i, UniCel DxC 680i, and UniCel DxC 660i systems.

DateJun 9, 2014 (initiated May 14, 2014)
StatusTerminated
ReasonBeckman Coulter is initiating a field action because some Ultrasonic Transducer Kits were potentially assembled with incorrect mounting screws.
Root causeMixed-up of materials/components
ActionBeckman Coulter sent an "URGENT MEDICAL DEVICE RECALL" letter dated May 8, 2014 via first class mail. The letter identified the reason for recall, product information, issue, impact, action, resolution, contact information, and customer response form. Customers are advised to review historical quality control (QC) data and System Check results, and determine whether a review of patient test results is warranted. For questions regarding this letter, please contact our Customer Support Center: From our website: http://www.beckmancoulter.com/customersupport/support, By phone: call 1-800-854-36
Quantity82 units
DistributionWorldwide Distribution and USA Nationwide - and the countries of Australia, Canada, China, France, Hong Kong, Italy, Japan, Mexico, Portugal, Spain, Switzerland, Taiwan,and Turkey.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K042291UNICEL DXC 600 AND 800 SYNCHRON SYSTEMS
Beckman Coulter, Inc., Nov 12, 2004