ClearanceBrief
Recall Z-1680-2014 ยท OEI Drainage Catheter With Antibiotic

Bacterin International, Inc.

Wound Drain - Label reads in part "Elutia ***Hubless Coated Silicone Surgical Wound Drain***" Two different sizes, 10 units per box: 19 Fr Round Hubless part number 7110; 24 Fr Round Hubless part number 7118. AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system.

DateJun 2, 2014 (initiated May 15, 2014)
StatusTerminated
ReasonThe recalling firm reported there was inadvertent omission of quarterly (Q4 - 2013 and Q1 - 2014) dose audits/verifications on the affected products. All of the affected products were appropriately sterilized
Root causeUse error
ActionBacterin sent e-mail and FedEx letter with return receipt request to it's single consignee on May 15, 2014. Consignee is asked to separate the recalled product in a secure location for return to Bacterin International, Inc. Consignee is asked to complete tracking/verification form included in the letter/e-mail and call 1-888-886-9354 from 8:00AM to 5:00PM (Mountain Time) or send an e-mail to CS@Bacterin.com to obtain a Return Authorization (RA) number.
Quantity390
DistributionDistributed to one foreign consignee in Lebanon.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K063245ELUTIA COATED CLOSED SURGICAL WOUND DRAIN
Bacterin International, Inc., Nov 28, 2007