Recall Z-1673-2014 ยท FRN Pump, Infusion

Bayer Corp

Medrad Continuum MR Infusion system- non-wireless system The MEDRAD Continuum MR Infusion System is designed for patients who require medications and other fluids during an MR procedure. It is intended to provide infusion therapy directly prior to, during and immediately after the MR procedure functioning while either stationary or mobile. It is not intended to provide long-term patient care outs

DateMay 30, 2014 (initiated Apr 3, 2014)
StatusTerminated
ReasonThere is a potential safety risk associated with potentially damaged locking pins within the bracket of non-wireless versions of Continuum systems shipped prior to March 2008.
Root causeComponent design/selection
ActionBayer sent an Urgent Medical Device Recall and Removal Letters dated April 3, 2014, and April 30, 2014, were sent to direct accounts to notify them of the bracket issue and to provide additional instructions for testing and/or returning the brackets. As a critical medical needs site currently using Continuum with a recalled bracket you have two options: 1. Immediately return your Continuum(s) for refund. 2. Take actions to ensure continued safe use of your system associated with the bracket recall. You will still be eligible for a refund when you return the system no later than June 30, 2015.
Quantity1347
DistributionWorldwide Distribution - USA (nationwide) including the countries of Australia, Brazil, Belgium, Canada, Germany, Singapore, South Africa, Poland, Mexico and UAE.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K073051MEDRAD CONTINUUM MR INFUSION SYSTEM WITH WIRELESS REMOTE ACCESSORY
Medrad, Inc., May 30, 2008