Recall Z-2544-2014 ยท OYC Pump, Infusion, Insulin, To Be Used With Invasive Glucose Sensor
Medtronic MiniMed Inc.
Medtronic MiniMed Paradigm Insulin Infusion Pumps, Paradigm models: MMT-511, MMT-512, MMT-712, MMT-712E, MMT-515, MMT-715, MMT-522, MMT-522K, MMT-722, MMT-722K, MMT-523, MMT-523K, MMT-723, MMT-723K, MMT-554, and MMT-754. MiniMed 530G: MMT-551, and MMT-751 Paradigm insulin infusion pumps are intended for use to deliver basal and bolus insulin to persons with diabetes.
| Date | Aug 22, 2014 (initiated Mar 13, 2014) |
| Status | Terminated |
| Reason | Medtronic is recalling the Medtronic MiniMed Paradigm Insulin Infusion Pumps because they have received reports that users have accidentally programmed the pump to deliver the maximum bolus amount. |
| Root cause | Device Design |
| Action | The firm, Medtronic, sent an "IMPORTANT MEDICAL DEVICE SAFETY INFORMATION" letter dated March 13, 2014 to its customers. The letter described the product, problem and actions to be taken. The letter informs the customers that because accidental button pressing errors may occur it is important that customers always confirm the insulin dose flashing on the display is correct before pressing ACT to start delivery. The letter informs the customers of what safety settings will minimize the risk of an accidental programming error and what actions to be taken. Should you have any questions about |
| Quantity | 559,374 units total (444,374 units in US) |
| Distribution | Worldwide Distribution-Nationwide in US |
| FDA record | Open on fda.gov |