ClearanceBrief
Recall Z-1734-2014 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Ortho-Clinical Diagnostics

VITROS 4600 Chemistry System, Product Code/Catalog Number 6802445 for in vitro diagnostic use.

DateJun 9, 2014 (initiated Apr 10, 2014)
StatusTerminated
ReasonPotential for biased results when a Multiple Window (MW) code is associated with results generated by the VITROS 4600 Chemistry System, the VITROS 5600 Integrated System and the VITROS 5,1 Chemistry System. MW codes may be associated with results from any of the following VITROS Chemistry Products using multiple point (multipoint) rate measurements. The MicroSlide assays that could be affected
Root causeSoftware design
ActionImportant Product Correction Letters (CL2014-116, dated 4/10/2014) were sent on 4/10/2014 via FedEx overnight courier or ORTHO PLUS e-Communications to all customers with VITROS 5,1 FS, 5,1 FS Refurbished, 4600,and/or 5600 Systems. They were informed of the issue and advised to: -- Review VITROS ALT, ALTJ, ALKP, ALKPJ, AST, ASTJ, CHE, CHEJ, CK, CKJ, GGT, GGTJ, LDH and LDHJ results generated on VITROS 4600/5600 and 5,1 FS Systems. Do not report any results associated with an MW code. -- Perform repeat testing in singlicate on all samples with results having an MW code. Only results without a
QuantityDomestic - 82, Foreign - 213 units
DistributionWorldwide distribution including US Nationwide, Puerto Rico, Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain, and Venezuela.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K031924VITROS 5,1 FS CHEMISTRY SYSTEM, VITROS CHEMISTRY PRODUCTS DHDL REAGENT PACK, CALIBRATOR KIT 19, FS CALIBRATOR 1
Ortho-Clinical Diagnostics, Aug 7, 2003