ClearanceBrief
Recall Z-1847-2014 ยท NEP System/Device, Pharmacy Compounding

Baxter Corporation Englewood

Abacus 3.1, Pharmacy Calculator Product Usage: Provide pharmacy calculations for automated compounding.

DateJun 22, 2014 (initiated May 28, 2014)
StatusTerminated
ReasonBaxter Corporation is initiating a field correction following an upgrade to ABACUS Software v3.1, when Clinimix 4.25110 is used in an order, the Nutritional Summary and any labels containing energy information may display 1,000 times less protein than is actually present in the solution.
Root causeSoftware design
ActionBaxter sent an Urgent Device Correction letter dated May 27, 2014 via certified mail to affected customers. The letter identified the affected product, problem and actios to be taken. Customers were instructed to: 1. Review and distribute this communication to pharmacy staff and compounding staff that use ABACUS software. This notice is intended only for customers that have upgraded to version 3.1 of the software. 2. Please acknowledge receipt of this notification by completing the enclosed customer reply form and return it to Baxter by either faxing it to 1-224-270-5457 or scanning and e
Quantity42
DistributionUS Nationwide Distribution and the state of VA.
FDA recordOpen on fda.gov