Recall Z-1948-2014 ยท BTM Ventilator, Emergency, Manual (Resuscitator)

Ventlab LLC

Manual Resuscitator Bag Series: AF1000, AF2000, AF3000, AF4000, AF5000, BT2000, BT3000, BT4000, BT5000, BVM700, CPRM1000, CPRM2000, CPRM3000, PRO-1900, PRO-1000, PRO-2000, SC7000, SC8000, SC8020, SC8120, SC9000, SS3200, VN2000, VN3000, VN4000, VN5000, VT1000 , VN2102, VN2002. Emergency manual resuscitator for the Pediatric Population.

DateJul 3, 2014 (initiated May 16, 2014)
StatusTerminated
ReasonThe pop-off valve in the defective devices remain open and a squeeze of the bag may not generate enough force to force the duck bill valve open and therefore no air will be delivered to the victim.
Root causeProcess change control
ActionThe firm, Ventlab, issued a press dated May 14, 2014, stating Ventlab, LLC has initiated a voluntary medical device removal of a limited number of Ventlab" Resuscitator Bags. The bag series and lot numbers were included in a table as well as manufacture dates. In addition, Ventlab, LLC sent an "URGENT: MEDICAL DEVICE PRODUCT REMOVAL IMMEDIATE ACTION REQUIRED" letter dated May 19, 2014 to its distributors and customers. The letter described the product, problem and actions to be taken. The customers were instructed to review and examine their inventory for affected product; stop using them and
Quantity251,384
DistributionUS Nationwide Distribution in the states of OH, NY, TX, TN, FL, CT, and VA,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K012842V*CARE MANUAL RESUSCITATOR
Ventlab Corp., Sep 18, 2001
K953546MANUAL PULMONARY RESUSICITATOR
Medical Marketing Concepts, Feb 29, 1996
K982215MANUAL RESUSCITATOR WITH ACE
Medical Marketing Concepts, Sep 18, 1998