Recall Z-1634-2014 ยท DRY Monitor, Blood-Gas, On-Line, Cardiopulmonary Bypass

Terumo Cardiovascular Systems Corporation

CDI Blood Parameter Monitoring System; Catalog Numbers or Part Numbers - 145891, 500AVHCT; Serial Numbers - BPM SN B015552 (installed into CDI 500 Monitor SN 3535)

DateMay 19, 2014 (initiated May 9, 2014)
StatusTerminated
ReasonThere is a potential for inaccuracies in blood gas readings/measurements. There is a potential risk of respiratory alkalosis or cerebral vasoconstriction with low paCO2. If the Hemoglobin and Hematocrit measurements are not accurate, there is a potential risk of unnecesary blood transfusion with resulting risk of infection or inflammatory reaction. BPM shipped without meeting manufacturing accep
Root causeRelease of Material/Component prior to receiving test results
ActionOn 5/9/14, TCVS placed a phone call to the one affected customer regarding the recall. The phone call explained the reason for the recall, identification of affected product, potential hazard, and instructions for return of affected product.
Quantity1
Distributionone customer in IA
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K9729623M CDI BLOOD PARAMETER MONITORING SYSTEM 500
3M Company, Nov 6, 1997