Biomet, Inc.

EZPass Passer, Sterile single use arthroscopic instrument utilized in conjunction with a nitinol wire to aid in passing suture through soft tissue. Part numbers 904050, 904051, 904052, 904053, 904054, 904055, and 904056.

DateJun 16, 2014 (initiated Apr 15, 2014)
StatusTerminated
ReasonThe nitinol wire or monofilament may buckle within the EZPass Suture Passer and become kinked making the wire/monofilament difficult to use.
Root causeDevice Design
ActionOn 2/28/14, customers received a letter stating these devices have been recalled following an investigation which identified that the nitinol wire or monofilament may buckle within the EzPass Suture Passer and become kinked making the wire/monofilament difficult to use. The letter also stated, there is no adverse health outcome expected for the patient and this action requires the immediate location and discontinued use of the product and its return to Biomet.
Quantity102
DistributionNationwide Distribution.