Recall Z-1706-2014 ยท KPY Shield, Protective, Personnel
GE Healthcare, LLC
Kenex Radiation Shield and surgical lamps installed with GE Healthcare Interventional, Surgery, CT and X-Ray Imaging systems. Kenex Suspension installed with Radiation shield and lamp with following model numbers: E03971AA;E03981AA;E30421AA;E30431AA;E30591CA;E30591HD;E30591HL;E30591EB;E30511P;E30591AB;E30591B;E30511Q;E30591C;E30591A;E30591RR;E30591CB;E30591E;E33591PA;E33591PC. Product Usage: Pe
| Date | Jun 3, 2014 (initiated May 2, 2014) |
| Status | Terminated |
| Reason | GE Healthcare has recently become aware of a potential safety issue due to the improper installation of a Kenex ceiling suspended radiation shield and surgical lamp. This issue is not related to the design of the Kenex product. The affected products are limited to installations completed or contracted by GE Healthcare. There has been a reported incident of a radiation shield falling from the ceil |
| Root cause | Equipment maintenance |
| Action | GE Healthcare sent an "Urgent Medical Device Correction" letter # 12209 dated May 2, 2014. The letter was addressed to Director of Biomedical Engineering, Director of Radiology and Chief of Cardiology. The letter described the Safety Issue, Safety Instructions, Affected product Details, Product Correction and Contact Information. For questions they should call 1-800-345-2700. For other countries, they should contact their local GE Healthcare Service Representative. . |
| Quantity | 3221 (108 US, 3113 OUS). |
| Distribution | Worldwide Distribution - US Nationwide, ALBANIA, ALGERIA, ANGOLA, ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BAHRAIN, BELARUS, BELGIUM, BOLIVIA, BOSNIA & HERZEGOVINA, BRAZIL, BULGARIA, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, CROATIA, CZECH REPUBLIC, DENMARK, DOMINICAN REPUBLIC, ECUADOR, EGYPT, F |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.