Recall Z-1806-2014 ยท MUJ System, Planning, Radiation Therapy Treatment

RAYSEARCH LABORATORIES AB

RayStation software version 4.0.3.4 RayStation is a software system designed for treatment planning and analysis of radiation therapy.

DateJun 17, 2014 (initiated Apr 25, 2014)
StatusTerminated
ReasonThis notice concerns a problem with the computation of ROI voxel volumes for cases where the CT has variable slice spacing. The dose grid volumetric representation of an ROI can be wrong. This affects all dose-volume properties for such ROIs, including DVH, dose statistics, clinical goals and constraints or objective functions. Furthermore, if material override ROIs are used, or if the external is not represented by contours, the bug can trigger an error in the 3D dose distribution.
Root causeSoftware design
ActionOn April 25, 2014, RaySearch emailed all consignees of the affected product a copy of the recall notice as well as the response forms. Back up notification methods include phone calls and site visits.
Quantity25 sites/245 units (US), 25 sites/124 units (foreign)
DistributionWorldwide Distribution-USA (nationwide) including the states of AZ, CA, FL, IL, MA, MD, MO, MS, NJ, NV, NY, OH, PA, TN, VA, and WA, and the countries of Australia, Belgium, Germany, France, Italy, New Zealand, Republic of Korea, Switzerland, Spain, and United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K130617RAYSTATION
Raysearch Laboratories AB, May 10, 2013