ClearanceBrief
Recall Z-1615-2014 ยท KWI Prosthesis, Elbow, Hemi-, Radial, Polymer

Biomet, Inc.

ExploR 7x26mm Modular Radial Stem, Catalog number 11-210062. Elbow prosthesis orthopedic implant.

DateMay 13, 2014 (initiated Apr 22, 2014)
StatusTerminated
ReasonThis lot of the ExploR Modular Radial Stem may be missing the wedge ramp threads. If the wedge ramp threads are missing from the stem, the likelihood of the set screw loosening from the radial stem intraoperatively is increased. If the set screw loosens, a revision surgery may be necessary.
Root causeProcess control
ActionBiomet sent an URGENT MEDICAL DEVICE SAFETY ALERT dated April 22, 2014 to affected customers. The letter identified the affected product, problem and actions to be taken. Questions related to this notice should be directed to (574) 372-1570, Monday through Friday, 8 a.m. to 5 p.m.
Quantity8 devices
DistributionUS Nationwide Distribution in the states of NY, MD, NC, SC, MI, OR, and NV.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K040611MODULAR RADIAL HEAD REPLACEMENT DEVICE
Biomet, Inc., Jun 4, 2004
K051385EXPLOR MODULAR RADIAL HEAD REPLACEMENT DEVICE
Biomet, Inc., Aug 5, 2005