ClearanceBrief
Recall Z-1624-2014 ยท GBX Catheter, Irrigation

Hospira Inc.

LATEX-FREE; T-U-R Y-Set; Nonvented Set for Transurethral Resection Procedures; 96 inch (244 cm); For Flexible Irrigation Container Systems; List No. 6543-01; single use; Hospira Product Usage: For use in transurethral resection procedures

DateMay 16, 2014 (initiated Mar 25, 2014)
StatusTerminated
ReasonThe Tyvek cover may not be completely sealed to the rest of the package allowing openings at the edge of the sterile package for the following products: List No. 04693-01, Latex Free Extension Set; List No. 06543-01, Nonvented Y-Set for Transurethral Resection; List No. 06544-01 Cystoscopy / Irrigation Set; List No. 06599-01, Large Bore Y-Irrigation Set (Latex-Free); and, List No. 15239-01, 4-Lea
Root causeProcess design
ActionHospira sent an URGENT MEDICAL DEVICE RECALL letter dated March 25, 2014 to all affected customers. The letter included instructions for the direct accounts to: 1) check their inventories and quarantine any of the affected products; 2) complete and return the attached reply form to the fax number or e-mail address on the form; 3) inform potential users of these products of this recall; 4) return the affected products to Stericycle using the labels provided with the letter (additional labels or reply forms can be obtained by calling Stericycle at 877-377-5128, Monday - Friday, 8:00 AM - 5:00 PM
Quantity1,109,321 set (US); 15,113 sets (Foreign)
DistributionWorldwide Distribution - US: Nationwide in the states of (AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, UT, WA, WI, WV, WY) including District of Columbia and Puerto
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.