ClearanceBrief
Recall Z-1550-2014 ยท FMT Warmer, Infant Radiant

GE Healthcare

Panda Wall-Mount Infant Warmers, integrated with Nellcor SpO2. Controlled infrared heat to neonates who are unable to thermo-regulate based on their own physiology with integrated monitoring feature used for continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by a SpO2 sensor).

DateMay 6, 2014 (initiated Apr 2, 2014)
StatusTerminated
ReasonMedical device software may be defective and may cause user-set alarm limits to result in false positive or false negative Oxygen saturation and pulse rate alarm notifications to the caregiver.
Root causeSoftware design
ActionGE Healthcare notified Biomedical/Clinical Engineering managers, Labor & Delivery/ NICU Nurse managers, Respiratory Therapy Managers and Directors of Risk Management by letter 04/02/2014. End users were advised that they may continue to use the system provided they follow the GE recommended actions: 1) If in use with a patient, turn off the Nellcor SpO2 function in the control panel and switch to an approved alternate form of SpO2 measurement. 2) Prior to using the affected GE warmer with the Nellcor SpO2 function, or resuming the use of this function with any patient, contact your hospital Bi
Quantity15 units
DistributionWorldwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K122267GIRAFFE AND PANDA WARMERS
Ohmeda Medical, A Division of Datex-Ohmeda, Inc., Oct 2, 2012
K123309GIRAFFE AND PANDA WARMERS
Ohmeda Medical, Feb 5, 2013