Recall Z-1618-2014 ยท KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System

Fresenius Medical Care Holdings, Inc.

Fresenius 2008T Hemodialysis Machine with Revision K Actuator Test Boards and also sold as replacement parts for use with Fresenius 2008-Series Hemodialysis Machines (2008K, 2008K2, 2008K@home and 2008T). Part Number: 190234 Revision K

DateMay 13, 2014 (initiated Apr 25, 2014)
StatusTerminated
Reason2008-Series Hemodialysis Machines with Revision K Actuator Test Boards may fail to detect a disconnected Ultrafiltration (UF) Pump
Root causeDevice Design
ActionFresenius Medical issued Urgent Recall letter issued on 4/25/14 by certified mail with signature confirmation and fax-back form. Customers instructed to examine their inventory to determine whether they have any of the affected 2008T machines or if they have installed any of the affected replacement boards. If customers have the affected machines or replacement boards, they will be instructed to contact FMC-RTG Technical Service to have the Actuator Test Boards replaced. Contact the FMCNA Technical Service Team at 1-800-227-2572.
Quantity79 units
DistributionUSA (nationwide) and the country of Canada.*
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K121341FRESENIUS 2008T HEMODIALYSIS MACHINE WITH BIBAG SYSTEM