ClearanceBrief
Recall Z-2735-2014 ยท PAM Gram-Positive Bacteria And Their Resistance Markers

Nanosphere, Inc.

Verigene BC-GP Kit (Catalog number 20-005-018); 1) 20 Verigene BC-GP Test Cartridges; are qualitative multiplexed in vitro diagnostic tests for simultaneous detection and identification of selected gram-positive (BC-GP) and gram-negative (BC-GN) bacteria and resistance markers.

DateSep 30, 2014 (initiated Jun 17, 2013)
StatusTerminated
ReasonThere is a specific Extraction Tray lot containing Tips that may slightly increase the occurrence of Process SP Processing Error when used due to the Processor SP experiencing difficulty releasing the Tip to the final location in the Tip Assembly. The difficult releasing the Tip can result in a Processing Error and occasionally the Tip to be dropped into the Processor SP.
Root causeProcess change control
ActionUndated recall notification letters were sent to all direct accounts (customers) June 17 - 19, 2013. The letters included instructions for customers to: 1) check their inventory for the recalled products; 2) set any recalled products aside for return; and, 3) contact Nanosphere to obtain replacement tip holders. Customers with questions about this recall can contact Nanosphere via e-mail at technicalsupport@nanosphere.us or by calling 888-837-4436.
Quantity1,660 extraction trays
DistributionWorldwide Distribution:US (Nationwide) including states of: AZ, CA, FL, IL, KY, MD, MN, MT, OH, PA, TN, TX, VA, WA, WI; and countries of: Belgium, China, and Saudi Arabia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K122514VERIGENE GRAM-POSITIVE BLOOD CULTURE (GC-GP) NUCLEIC ACID TEST
Nanosphere, Inc., Sep 19, 2012