Recall Z-0131-2015 ยท CHI Lipase-Esterase, Enzymatic, Photometric, Lipase

Abbott Laboratories, Inc

Lipase Reagent, 7D80. Used to measure lipase in human serum on the Architect system.

DateOct 28, 2014 (initiated Apr 7, 2014)
StatusTerminated
ReasonPackage inserts have incorrect SmartWash parameters for Triglyceride.
Root causeLabeling Change Control
ActionFirm issued a Product Correction Letter to all Lipase reagent customers who received lots 43827UQ12, 43819UQ03, 44449UQ05, 44127UQ06, 44334UQ07, 43994UQ07, 44596UQ08, 44775UQ09, and 44930UQ11. The letter identified the affected product and provided an explanation of the problem, as well as, the patient impact and actions to be taken. The Product Correction Letter provided guidance for the laboratory to correctly configure the Architect cSystem's Lipase assay parameters. Customers are to retain a copy of the letter for their lab records, and forward a copy to any laboratories that the produc
Quantity11916 units
DistributionWorldwide Distribution: US (nationwide) and countries of: Algeria, Angola, Argentina, Armenia, Australia, Austria, Bangladesh, Belgium, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Columbia, Croatia, Czech Republic, Dominican Republic, Egypt, Estonia, France, Germany, Greece, Guat
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.