Recall Z-1721-2014 ยท JAK System, X-Ray, Tomography, Computed

Siemens Medical Solutions USA, Inc

Siemens SOMATOM Emotion 16, Emotion 6, and SOMATOM Spirit Computed Tomography Systems Product Usage: To produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes taken at different angles.

DateJun 4, 2014 (initiated Mar 21, 2014)
StatusTerminated
ReasonThe rubber damper may degrade and break, causing loosening of the motor mounting plates, which could lead to a series of events that could place the operator and patient at risk of injury.
Root causeComponent design/selection
ActionSiemens sent a Customer Safety Advisory Notice/ Field Safety Notice letter dated March 21, 2014, to affected customers. The letter identified the affected product, problem and actions to be taken. The letter informed customers that the system modification would be implemented by Siemens service specialists in April 2014 to correct the problem.
Quantity345
DistributionUS Nationwide Distribution in the states of IA, TX, FL, NE, NY, OH, NC, KS, IL, CA, OK including PR.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K023687SOMATOM EMOTION 6 CT SYSTEMS
Siemens Medical Systems, Inc., Nov 22, 2002
K042328SOMATOM SPIRIT
Siemens Medical Systems, Inc., Sep 14, 2004
K050297SOMATOM EMOTION 16
Siemens Medical Solution, Inc. USA, Mar 1, 2005