ClearanceBrief
Recall Z-1487-2014 ยท IYN System, Imaging, Pulsed Doppler, Ultrasonic

Boston Scientific Corporation

iLab Ultrasound Imaging System; iLab 3.0 (Polaris 1.0) software. Used for ultrasound examinations of intravascular pathology.

DateApr 22, 2014 (initiated Apr 3, 2014)
StatusTerminated
ReasonSpecific iLab Ultrasound Imaging Systems using newest 3.0 (Polaris 1.0) software change the pullback speed from the default setting of 0.5 mm/sec to 1.0 mm/sec, the Longview distance measurement may display an inaccurate value that is double the actual distance.
Root causeNonconforming Material/Component
ActionBoston Scientific sent an Urgent Medical Device Field Correction letter to all affected customers via hand delivery by BSC sales representatives. The letter identifed the affected product, problem and actions to be taken. Customers were asked to immediately post the Field Correction information in a visible location near the system(s) to ensure it is easily accessible to all users of the device. For question contact your local Sales Representative.
Quantity9 units
DistributionUSA Nationwide Distribution in the states of NY, NC, DC, FL, PA
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.