Recall Z-1642-2014 ยท MUE Full Field Digital, System, X-Ray, Mammographic

Fujifilm Medical System USA, Inc.

MXR-35 Diagnostic X-ray High Voltage Generator manufactured by DRGEM Corporation. x-ray generator

DateMay 21, 2014 (initiated Jun 29, 2012)
StatusTerminated
ReasonThe MXR-35 X-ray Generator's labeling did not include the certification statement as required by 21 CFR Part 10 10.2.
Root causeRadiation Control for Health and Safety Act
ActionFujifilm will replace the violative labels with replacement labels provided by DRGEM, to bring the X-ray generator into compliance. You stated that you expect for all nine installed units to be corrected by the end of May 2014. The seven units in the Stamford, CT warehouse will be corrected when the generators are delivered to customer sites. Under 21 CFR 1003.31(c)(2), this exemption is granted subject to the following conditions: 1. You must correct the missing certification statement for all units shipped into the US within the timeline you specified, by the end of May 2014. 2. You must e
Quantity14
DistributionWorldwide Distribution - US Distribution including the states of TX and KY and Internationally to China.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.