ClearanceBrief
Recall Z-1654-2014 ยท JLW Radioimmunoassay, Thyroid-Stimulating Hormone

Siemens Healthcare Diagnostics, Inc

ADVIA Centaur Systems TSH3 Ultra , Ready Pack, 500 test Catalog Number: 06491080; Siemens Material Number: 10282379. Intended for the invitro diagnostic use in quantitative determination of various assays on the ADVIA Centaur and ADVIA Centaur XP analyzers.

DateMay 23, 2014 (initiated Apr 10, 2014)
StatusTerminated
ReasonSamples containing fluorescein may show interference with the ADVIA Centaur Systems TSH3 Ultra, Vitamin D and the ADVIA Centaur BRAHMS Procalcitonin assays.
Root causeDevice Design
ActionSiemens Healthcare Urgent Medical Device Correction (UMDC) was mailed to all affected Siemens Healthcare Diagnostics customers in the United States on April 11, 2014 for customer receipt on April 14, 2014. This notice informed the customers of fluorescein interference with the ADVIA Centaur Systems TSH3-Ultra and Vitamin D Total assays. This notice instructs customers that the TSH3-Ultra and Vitamin D assays can continue to be used for testing any sample that does not contain fluorescein. However, testing blood drawn from a patient, without allowing clearance of fluorescein is not recommended.
Quantity155,345 kits
DistributionWorldwide Distribution-USA (nationwide) and the countries of Afghanistan Algeria Armenia Austria Bahrain Belgium Bulgaria Burkina Faso Canary Islands Croatia Cyprus Czech Republic Denmark Estonia Finland France Germany Great Britain Greece Hungary Iceland Iran Iraq Ireland Israel Italy Jordan Kazak
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K083844ADVIA CENTAUR TSH3-ULTRA METHOD
Siemens Healthcare Diagnostics, May 8, 2009