Recall Z-1480-2014 ยท LNH System, Nuclear Magnetic Resonance Imaging

GE Healthcare, LLC

GE Healthcare, Discovery MR450, Discovery MR750, Optima MR450w, Optima MR 450w GEM, Discovery MR750w, and Discovery MR750w GEM products. Magnetic Resonance Diagnostic Device.

DateApr 21, 2014 (initiated Apr 2, 2014)
StatusTerminated
ReasonPotential safety issue with gradient coil electromechanical connections associated with GE MR Products. Gradient coil electromechanical connections may loosen causing an increase in resistance. The increased resistance can lead to increasing heat during scanning which, in turn, can cause material in the Rear Endbell enclosure to overheat, generate smoke in the scan room, and cause some localize
Root causeProcess change control
ActionConsignees were sent on 4/2/2014 a GE Healthcare "Urgent Medical Device Correction" letter dated April 2, 2014. The letter was addressed to Hospital Administrators / Risk Mangers, Radiology Department Mangers and Radiologists. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product correction and Contact Information (Fore questions, contact your local service representative ).
Quantity1,546 (704 US, 842 OUS).
DistributionWorldwide Distribution-US (nationwide) except the states of ND and RI, including DC, GU, PR, and the countries of ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA,AZERBAIJAN, BAHRAIN, BELARUS, BELGIUM, BRAZIL, CANADA, CAYMAN ISLANDS, CHILE, CHINA, CZECH REPUBLIC, DENMARK, EGYPT, FINLAND, FRANCE, FRENCH POLYNE
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K081028SIGNA 3.0T MR750 SYSTEM
Ge Medical Systems, LLC, Apr 25, 2008
K083147DISCOVERY MR450
Ge Medical Systems, LLC, Nov 4, 2008
K091536OPTIMA MR450W
Ge Medical Systems, LLC, Jul 17, 2009
K103327DISCOVERY MR750W 3.0T
Ge Healthcare Japan Corporation, Sep 30, 2011
K103335GEM OPTION FOR 1.5 MRI SYSTEMS
GE Healthcare, Mar 7, 2011