Recall Z-1462-2014 ยท OWB Interventional Fluoroscopic X-Ray System

GE OEC Medical Systems, Inc

The OEC 9900 Elite mobile fluoroscopy system (image intensified fluoroscopic x-ray) is designed to provide fluoroscopic and spot-film images of the patient during diagnostic surgical and interventional procedure. Examples of clinical application may include cholangiography, endoscopy, urologic, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedure. The system may

DateApr 16, 2014 (initiated Mar 31, 2014)
StatusTerminated
ReasonGE Healthcare has identified a potential safety issue related to the screws that attach the workstation monitor spring arm to the extension arm on certain OEC 9900 systems, causing the spring arm and monitor to fall forward.
Root causeDevice Design
ActionGE Healthcare sent an Urgent Medical Device Correction letter dated March 31, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Use caution when positioning or extending the workstation articulating arm. In addition, refrain from positioning staff or patients directly underneath or in front of a workstation monitor. Your local GE Healthcare Field Engineer will contact you to coordinate the replacement of the screws, and ensure that your system is operating safely and meets all specifications. These activities wil
Quantity453 units
DistributionWorldwide Distribution - USA, and the countries of Australia, Argentina, Canada, China, Cambodia, Columbia, Japan, Ecuador, Japan, Korea, Hong Kong, India, Israel, Libya, Mexico, Nigeria, Peru, Russia, Thailand, Taiwan, Saudi Arabia and Singapore.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K0619539900 PLUS MOBILE FLUOROSCOPY SYSTEM WITH 3D AND NAVIGATION OPTIONS
Ge Oec Medical Systems, Aug 15, 2006
K073543OEC 9900 ELITE
Ge Oec Medical Systems, May 2, 2008
K082781OEC 9900 ELITE
Ge Healthcare Surgery, May 1, 2009