Recall Z-1461-2014 ยท JAK System, X-Ray, Tomography, Computed

Philips Medical Systems (Cleveland) Inc

Brilliance CT 16, Brilliance CT 64, Brilliance CT Big Bore Oncology, Brilliance CT Big Bore Radiology, Brilliance iCT, Ingenuity CT, Ingenuity Core, Ingenuity Core 128 and Ingenuity Flex scanners are Computed Tomography X-ray systems.

DateApr 16, 2014 (initiated Mar 27, 2014)
StatusTerminated
ReasonBlock assembly-lock stop subframe of service latch was not manufactured to engineering specifications from the supplier, causing premature fracture of the threaded rod on the patient support service latch. As a result, the table top can become free floating due to the disengaged service latch in the table tops subframe.
Root causeNonconforming Material/Component
ActionA Field Safety Notice was sent to all US consignees via certified mail. Market representatives will distribute the letter internationally. Additionally, a Field Service Engineer will contact each site to schedule a time to implement the Field Change Order.
Quantity135
DistributionNationwide Distribution - USA including the states of CA, FL, IL, IN, KY, MD,SC, and TX.
FDA recordOpen on fda.gov