Recall Z-1476-2014 ยท LIB Device, General Purpose, Microbiology, Diagnostic

BioMerieux SA

FB Reagent (REF 70562). FB reagent is an additional test used for revealing the results of some miniaturized biochemical test included in the API and Rapid ID 32 strips. The API and ID 32 strips product lines are a standardized system combining several biochemical tests, which enables group or species identifications of microorganisms.

DateApr 18, 2014 (initiated Mar 6, 2014)
StatusTerminated
ReasonbioMerieux identified a visual defect and activity issue on the ZYM B reagent (REF 70493) and FB reagent (REF 70562). The activity issue lead to a false negative result of some Rapid ID 32 strips and API biochemical tests using the ZYM B reagent or FB reagent
Root causeOther
ActionLetters were issued to customers on March 10, 2014. The letter contained instructions to be used if they still were using the product. The letter also included a report form which was to be returned.
Quantity33,853
DistributionWorldwide Distribution-USA (nationwide) including DC and Puerto Rico, and the states of AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, VA, WV, WA, WI, and WY and the