ClearanceBrief
Recall Z-1324-2014 ยท GCJ Laparoscope, General & Plastic Surgery

Teleflex Medical

10mm Weck Vista Cannula-only, Catalog No. 405910C, 12mm Weck Vista Cannula-only, Catalog No. 405912C. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals.

DateApr 2, 2014 (initiated Mar 11, 2014)
StatusTerminated
ReasonComplaints of leakage of insufflation gas through the device.
Root causeNonconforming Material/Component
ActionLetters were issued on March 11, 2014 to customers informing them of the recall. The letter asked the consignee to immediately discontinue use and quarantine any products on hand. The letter included a Recall Acknowledgement Form which is to be returned. A letter was also issued to distributors on March 11, 2014, requesting them to conduct a sub-recall.
Quantity55,521 total
DistributionWorldwide Distribution-USA (nationwide) including Puerto Rico and the states of AL, AZ, AR, CA, CO, CT, FL, GA, IL, IN, KS, KY, MD, MA, MI, MN, MS, MO, NH, NJ, NY, NC, ND, OH, OR, PA, SC, TN, UT, VA, WA, WV, WI and the countries of Australia, Brazil, Canada, China, Israel, Korea, Malaysia, Singapore
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K121796WECK REUSABLE OBTURATOR
Teleflexmedical, Inc., Oct 11, 2012