Recall Z-1468-2014 ยท JXG Shunt, Central Nervous System And Components

Codman & Shurtleff, Inc.

CODMAN CERTAS Programmable Valves, central nervous system shunt, models as follows: 82-8800 In Line Valve Only 82-8801 In Line Valve with Catheter and Accessories 82-8802 In Line Valve with Unitized Catheter and Accessories 82-8803 In line Valve with Unitized BACTISEAL Catheter and Accessories 82-8804 In Line Valve only with SIPHONGUARD Device 82-8805 In Line Valve with SIPHONGUARD Device, Cathe

DateApr 17, 2014 (initiated Dec 13, 2013)
StatusTerminated
ReasonEnsuring all customers have the tools to assist in verifying the setting of the valve on x-ray images and the location and orientation of implanted CODMAN CERTAS Programmable Valves.
Root causeProcess control
ActionCodman Neuro issued notification packages dated 12/13/13 to Domestic Hospitals and international accounts. Specific letters and associated inclusion of the additional aids, include A) a valve setting verification guide with instructions for use with x-ray images, and B) a plastic valve cut out overlay). A response reply form was included to be completed and returned to confirm effectiveness of the notification.
Quantity9,445 units
DistributionNationwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K112156CODMAN CERTAS PROGRAMMABLE VALVE SYSTEM, CODMAN CERTAS THERAPY MANAGEMENT SYSTEM
Medos International SARL, Oct 17, 2011
K113526CODMAN CERTAS PROGRAMMABLE VALVE SYSTEM
Medos International SARL, Feb 16, 2012