Recall Z-1470-2014 ยท KLT Enzyme Immunoassay, Carbamazepine

Ortho-Clinical Diagnostics

VITROS Chemistry Products CRBM Slides, REF 889 2382, IVD -- Ortho-Clinical Diagnostics, Inc. For in vitro diagnostic use only. VITROS Chemistry Products CRBM Slides quantitatively measure carbamazepine (CRBM) concentration in serum and plasma using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.

DateApr 17, 2014 (initiated Mar 5, 2014)
StatusTerminated
ReasonThe firm identified a potential for biased carbamazepine (CRBM) results to be generated when using VITROS CRBM Slides, Lot 3920-0080-8403.
Root causeRelease of Material/Component prior to receiving test results
ActionThe firm, Ortho Clinical Diagnostics, sent an "URGENT PRODUCT CORRECTION NOTIFICATION" letter (Ref. CL14-086; dated March 5, 2014) via FedEx Overnight Courier and/or US Postal Service Priority Mail (for PO Boxes only) to its US consignees (OCD Direct, Drop-Ship and US Federal Government). In addition, OCD proactively communicated with the 3 affected customers via phone to inform them of the potential issue and expedite product replacement. The letter described the product, problem and actions to be taken. The customers were instructed to immediately discontinue using and discard all remain
Quantity20 units
DistributionUS Distribution to states of: IN, MD, and TN.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K980283VITROS CHEMISTRY PRODUCTS CRBM SLIDES
Johnson & Johnson Clinical Diagnostics, Inc., Feb 25, 1998