Recall Z-1464-2014 ยท KPS System, Tomography, Computed, Emission

Philips Medical Systems (Cleveland) Inc

Ingenuity Core 128, Computed Tomography X-ray systems, Philips Medical Systems (Cleveland), Inc.

DateApr 16, 2014 (initiated Mar 6, 2014)
StatusTerminated
ReasonA customer reported that after upgrading to software version 3.5.5 from 3.5.4, reconstructions would intermittently not complete.
Root causeSoftware in the Use Environment
Action.The firm, Phillips, sent an " URGENT Medical Device Correction" letter dated 2013 March 06, to its costumers . The letter disc ribes the product, problem and actions to be taken. The costumers were instructed to Verify that all planned image reconstructions are completed, and Perform off-line reconstructions if expected images are missing. The firm will release A Field Change Order (FCO) to correct the issue with software version release 3.5.7.and A Philips Field Service Engineer will contact you to schedule the software installation at your site. If you need any further information or sup
Quantity116 units
DistributionWorldwide Distribution:US ( Nationwide) Including stages of MI, OR, PA, TX & WV; and Countries of: Australia, Bangladesh, Belgium, Brazil, Bulgaria, China, France, Germany, India, Indonesia, Ireland, Japan, Nicaragua, Oman, Palestine, Russia, South Korea, Spain, Switzerland, Thailand, Turkey & Unit