ClearanceBrief
Recall Z-1526-2014 ยท LGW Stimulator, Spinal-Cord, Totally Implanted For Pain Relief

Medtronic Neuromodulation

Medtronic, External neurostimulator, ENS (37021) & ENS (37022). Including the external neurostimulators which are associated with software application run on the NVision Clinician Programmer and used during trialing for spinal cord stimulation (SCS) patients.

DateApr 29, 2014 (initiated Feb 27, 2014)
StatusTerminated
ReasonMedtronic neurostimulators (used with pain and deep brain stimulation therapy) have a cycling feature that, when enabled, allows stimulation output to be toggled on and off based on the programmed settings. The use of this feature is determined by the physician and set using the clinician programmer. Current labeling indicates the use of cycling improves device longevity and recharge interval
Root causeLabeling False and Misleading
ActionMedtronic sent a Medical Device Correction letter dated February 2014 to all affected consignees. The letter identified the product, the problem, and the action to be taken by the consignee. Consignees were instructed to do the following: If you have patients with cycling enabled or wish to use cycling, use the updated calculations and graphs available online at professional.medtronic.com/cycling to assess the impact of cycling on device longevity. Medtronic will update the product labeling in the second half of 2014. " Remind your patients to continue to check their battery status. Instruct
Quantity170,286 US, 73,921 OUS for all devices.
DistributionNationwide Distribution - all states including DC and Puerto Rico.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.