Recall Z-1311-2014 ยท MKJ Automated External Defibrillators (Non-Wearable)

ZOLL Medical Corporation

Zoll Medical X Series, PropaqMD, PropaqM Defibrillator/Pacemaker/Monitor With System Software Version 02.10.02.00 or Higher, automated external defibrillator.

DateMar 31, 2014 (initiated Mar 5, 2014)
StatusTerminated
ReasonDefibrillator/Pacemaker/Monitor May Latch in a Continuous Device Reset Loop and cannot be used
Root causeSoftware design
ActionZoll Medical issued a Notification letter daetd March 5, 2014, to all affected consignees. The notification letter provides information on how to avoid the issue and instructs customers to contact ZOLL to get affected devices updated. Customers will be given the option to have a software upgrade kit sent to them to update their devices or return devices to ZOLL Service department in Chelmsford Massachusetts to have the device upgraded with the latest software release (Software version 02.16.04.00). Customers with questions were instructed to contact the 24/7 technical support numbers 1 (800)
Quantity3,169 units
DistributionWorldwide Distribution - USA (nationwide) and Internationally to Canada, AE, Afganistian, Australia, Belgium, Czech Republic , China, Chile, Croatia, Denmark, France, Great Britain, Hungary, Israel, Italy, Japan, Korea, Kuwait, Russia, Saudi Arabia, Singapore, South Africa,Thailand, and Turkey.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K112432ZOLL X SERIES
Zoll Medical Corporation, World Wide Headquarters, Mar 21, 2012
K112761ZOLL PROPAQ MD
Zoll Medical Corporation, World Wide Headquarters, Apr 25, 2012
K121367ZOLL PROPAQ XM
ZOLL Medical Corporation, Jun 21, 2012