Recall Z-1296-2014 ยท GAT Suture, Nonabsorbable, Synthetic, Polyethylene

Teleflex Medical

Sutures, Non-absorbable, Silk, Sterile, Rx only, Product Usage: Natural non-absorbable silk surgical suture is a non-absorbable, sterile, flexible multifilament thread composed of an organic protein called fibroin. This protein is derived from the domesticated species Bombyx mori (B. mori) of the family Bombycidae. Natural non-absorbable silk surgical suture is indicated for use in soft tissue

DateMar 26, 2014 (initiated Mar 11, 2014)
StatusTerminated
ReasonThe products are being recalled because they did not meet minimum needle attachment strength requirements.
Root causeEmployee error
ActionTeleflex sent an Urgent Medical Device Recall Notification letter dated March 11, 2014 to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to return all affected product to Teleflex Medical per the instructions provided in the letter. Customers were asked to complete the enclosed Recall Acknowledgment Form and fax it to 1-866-804-9881, Attn: Customer Service. For questions contact your local sales representative or Customer Service at 1-866-246-6990
QuantityTotal 32,271 ea.
DistributionWorldwide Distribution - U.S. Nationwide and the countries of Canada, Germany, and Ireland
FDA recordOpen on fda.gov