ClearanceBrief
Recall Z-1443-2014 ยท LGZ Warmer, Thermal, Infusion Fluid

Smiths Medical ASD, Inc.

Smiths Medical HOTLINE Fluid Warming Set L-70. Product Usage: intended for the delivery of warmed blood and intravenous fluids at normothermic temperatures under gravity conditions

DateApr 9, 2014 (initiated Mar 17, 2014)
StatusTerminated
ReasonInjection port of the Set may leak blood/ IV fluids from the septum of the injection port or the septum may become dislodged.
Root causeNonconforming Material/Component
ActionSmiths Medical issued Urgent Field Safety Notice Recall on 3/17/14 via mail service. Distributors will be instructed to notify their customers. The Field Safety Notice will also include an Urgent Recall Notice Confirmation Form (Confirmation Form) that consignees will be instructed to send back to Smiths Medical for carrying out the action and tracking effectiveness.Accounts are to inspect inventory and return recalled product. If you should have any questions regarding this information, please contact Shirley Duggan, Quality System Engineer at 1-781-763-9335 or by E-Mail Shirley.duggan@smiths
Quantity393,210
DistributionWorldwide Distribution - USA Nationwide in the states of: AU, CA, CL, CN, CZ, GT, ID, IN, IT, JP, LY, NZ, QA, SG, SK, TR, TW.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K911383HOTLINE FLUID WARMER
Level 1 Technologies, Inc., Oct 17, 1991