Recall Z-1558-2014 ยท GAM Suture, Absorbable, Synthetic, Polyglycolic Acid

Teleflex Medical

Fixt Suture Bonded Plus Violet Braided Polyglycolic Acid Coated Suture, Synthetic Absorbable Surgical Suture, Sterile, Rx Only, Teleflex Medical. Indicated for use in soft tissue approximation.

DateMay 7, 2014 (initiated Mar 11, 2014)
StatusTerminated
ReasonProduct does not meet minimum needle attachment strength requirements; thus, the needle can become detached from the suture during use.
Root causeEmployee error
ActionUrgent Medical Device Recall Notification letters were issued on March 11, 2014 to consignees. The letter requested that customers discontinue use and quarantine any product on hand. The attached Recall Acknowledgment Form should be completed and returned to Teleflex regardless of whether affected product is on hand or not. Once received a customer service representative will contact customers with instructions on how to return the affected product. Questions should be directed to Customer Service at 1-866-246-6990.
Quantity32,271 ea.
DistributionWorldwide Distribution -- USA, including the states of AL, CA, FL, GA, IL, IN, KS, LA, ME, MA, MI, MN, MS, MO, NV, NJ, NM, NY, OH, PA, RI, SC, TN, TX, UT, VA, WV, and WI; and countries of Canada, Germany, and Ireland.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K930738SURGICAL SUTURES, PGL, GUT, DEKLENE, OPTHAMEND ETC
Deknatel, Inc., Jul 26, 1994